Troubleshooting
Common fault cause analysis table |
||
Fault phenomenon |
Possible causes |
Removal method |
The actual flow rate of the electromagnetic flow meter is shown to be zero |
1、The actual flow direction of the fluid is opposite to the direction indicated by the sensor |
Change direction of sensor installation |
2、Electrode signal to ground short circuit |
Damage or break of signal cable insulation |
|
3、No excitation current |
The excitation cable is broken or damaged |
|
4、The electrode is completely covered by insulation |
Clean the electrode and remove the covering |
|
The actual flow of the electromagnetic flow meter is stable and the indication value is unstable |
1、The measured fluid is not filled with a measuring tube |
Change the position and manner of sensor installation so that the fluid is filled with the measuring tube |
2、The fluid under test contains more bubbles or oil |
Remove bubbles or oils |
|
3、Sensor ground fault |
Sensor grounding |
|
4、The electrode is heavily covered by insulation |
Cleaning electrode |
|
5、The power supply is unstable and the interference is large |
Separate power supply or anti interference stabilized power supply |
|
6、There is a strong interference device nearby |
Change installation location to avoid interference sources |
|
The pressure transmitter and the density sensor are not displayed or unstable |
1、Is the input voltage switched on? |
Reconnect voltage |
2、Input voltage positive and negative connection |
Changing voltage plus or Minus Voltage |
|
3、Sensor bad |
Replace sensor |
|
The pressure transmitter and the density sensor display very large and do not change |
1、 Sensor bad |
Replace sensor |
The pressure transmitter and the density sensor display unstable or almost zero |
1、 Is the sensor joint sealed? |
Re seal |